Diabetes Care
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Bernbaum, M.
Right arrow Articles by Mooradian, A. D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bernbaum, M.
Right arrow Articles by Mooradian, A. D.
Social Bookmarking
 Add to CiteULike   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?

Diabetes Care, Vol 16, Issue 10 1363-1366, Copyright © 1993 by American Diabetes Association


ARTICLES

Effectiveness of glucose monitoring systems modified for the visually impaired

M Bernbaum, SG Albert, S Brusca, J McGinnis, D Miller, JW Hoffmann and AD Mooradian
Department of Internal Medicine, St. Louis University School of Medicine, Missouri 63104.

OBJECTIVE--To compare three glucose meters modified for use by individuals with diabetes and visual impairment regarding accuracy, precision, and clinical reliability. RESEARCH DESIGN AND METHODS--Ten subjects with diabetes and visual impairment performed self-monitoring of blood glucose using each of the three commercially available blood glucose meters modified for visually impaired users (the AccuChek Freedom [Boehringer Mannheim, Indianapolis, IN], the Diascan SVM [Home Diagnostics, Eatontown, NJ], and the One Touch [Lifescan, Milpitas, CA]). The meters were independently evaluated by a laboratory technologist for precision and accuracy determinations. RESULTS--Only two meters were acceptable with regard to laboratory precision (coefficient of variation < 10%)--the Accuchek and the One Touch. The Accuchek and the One Touch did not differ significantly with regard to laboratory estimates of accuracy. A great discrepancy of the clinical reliability results was observed between these two meters. The Accuchek maintained a high degree of reliability (y = 0.99X + 0.44, r = 0.97, P = 0.001). The visually impaired subjects were unable to perform reliable testing using the One Touch system because of a lack of appropriate tactile landmarks and auditory signals. CONCLUSIONS--In addition to laboratory assessments of glucose meters, monitoring systems designed for the visually impaired must include adequate tactile and audible feedback features to allow for the acquisition and placement of appropriate blood samples.