Diabetes Care
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Olsen, C. L.
Right arrow Articles by Charles, M. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Olsen, C. L.
Right arrow Articles by Charles, M. A.
Social Bookmarking
 Add to CiteULike   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?

Diabetes Care, Vol 18, Issue 1 70-76, Copyright © 1995 by American Diabetes Association


ARTICLES

Diagnostic procedures for catheter malfunction in programmable implantable intraperitoneal insulin infusion devices

CL Olsen, DS Turner, M Iravani, K Waxman, JL Selam and MA Charles
Department of Medicine, College of Medicine, University of California, Irvine 92717.

OBJECTIVE--To evaluate the roles of 1) abdominal radiography, 2) a pressure diagnostic procedure (PDP) using a standardized diluent infusion into the catheter sideport, and 3) radiocontrast imaging of the catheter lumen as procedures for diagnosing catheter malfunction in diabetic patients implanted with a programmable intraperitoneal infusion device. RESEARCH DESIGN AND METHODS--Sixteen type I diabetic patients implanted with Infusaid programmable intraperitoneal insulin pumps were studied. The ability of the above three procedures to assist diagnosis of catheter malfunction and distinguish between occlusion and catheter breakage was retrospectively analyzed. Glycated hemoglobin was measured to determine the clinical importance of catheter malfunctions and decreases in pump flow due to insulin aggregation in the pump chamber. RESULTS--Mean glycated hemoglobin levels increased significantly from 8.0 +/- 0.3 to 9.0 +/- 0.4% (P < 0.05) before and after catheter malfunction, but not during pump flow slowdowns. Mean peak pressure during PDP was 1.96 +/- 0.14 psi (P < 0.01 vs. normal) in reversibly occluded catheters and 1.86 +/- 0.35 psi (P < 0.05 vs. normal) in broken catheters, compared with 1.32 +/- 0.23 psi in normal catheters. Decay times during PDP were > 50 s for both reversibly occluded and broken catheters (P < 0.001 vs. normal of 3.6 +/- 0.82 s). Abdominal radiographs and sideport injections of contrast material were used to distinguish the types of broken catheters. CONCLUSIONS--Catheter breakage and occlusion are complications in implantable insulin infusion systems and result in metabolic deterioration. The presence of a sideport allows pressure data and radiographic procedures to assist in determining the cause of catheter malfunction. A diagnostic algorithm was generated to improve efficiency in investigating catheter problems.