Diabetes Care
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Published online March 10, 2008
Diabetes Care 31:1120-1122, 2008
DOI: 10.2337/dc08-0145
© 2008 by the American Diabetes Association
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Online-Only Appendix
Right arrow All Versions of this Article:
dc08-0145v1
31/6/1120    most recent
Right arrow Purchase Article
Right arrow View Shopping Cart
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Google Scholar
Right arrow Articles by Hermanides, J.
Right arrow Articles by DeVries, J. H.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hermanides, J.
Right arrow Articles by DeVries, J. H.
Social Bookmarking
 Add to CiteULike   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?

Clinical Care/Education/Nutrition/Psychosocial Research
Original Research

No Apparent Local Effect of Insulin on Microdialysis Continuous Glucose- Monitoring Measurements

Jeroen Hermanides, MD, Iris M. Wentholt, MD, Augustinus A. Hart, PHD, Joost B. Hoekstra, MD, PHD and J. Hans DeVries, MD, PHD

From the Academic Medical Center, Amsterdam, the Netherlands

Corresponding author: Jeroen Hermanides, Academic Medical Center, P.O. Box 22660, 1100 DD Amsterdam, Netherlands. E-mail: j.hermanides{at}amc.uva.nl

OBJECTIVE—Data investigating the possible disturbing influence of insulin in the vicinity of continuous glucose monitoring (CGM) is lacking. We investigated the hypothesis that high local insulin concentrations would interfere with sensor readings.

RESEARCH DESIGN AND METHODS—Two microdialysis sensors were inserted in the periumbilical region of 10 continuous subcutaneous insulin infusion (CSII)-treated type 1 patients. A test sensor was inserted as close as possible to the insulin catheter and compared with a control sensor. Glucose peak and nadir were induced. Horizontal and vertical shifts were assessed using curve fitting, and mean absolute difference (MAD) between paired blood and sensor values were calculated.

RESULTS—Curve fitting showed no significant differences between the two sensors. MAD ± SD was 8.50 ± 3.47% for the test sensor and 9.21 ± 3.17% for the control sensor, P = 0.72.

CONCLUSIONS—Microdialysis CGM can be accurately performed in the proximity of CSII systems.

Abbreviations: CGM, continuous glucose monitoring • CSII, continuous subcutaneous insulin infusion • MAD, mean absolute difference